A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
|
Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
deflation: observed an opening through microscopic inspection assessed as fold flaw opening, observed broken device through microscopic inspection assessed as surgical damage, observed a missing piece of shell through microscopic inspection assessed as inconclusive and observed delamination of diaphragm valve through microscopic inspection assessed as adhesive failure.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases, wear abrasion and non penetrating nicks are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: a5, a6, d4, d9, e1, h3, h6.
|