The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: foreign objects in the nozzle, sticky connecting tube, sticky image guide protector and damaged on charged coupled device unit resulting in specified power to not be obtained.A root cause could not be identified.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: known inherent risk of device.Date of event is unknown.Additional information will be provided if available.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
|
The customer returned the gastrointestinal videoscope to the service center for a separate issue.During the device analysis, the investigation indicates that foreign objects in the nozzle, sticky connecting tube, sticky image guide protector and damaged on charged coupled device unit resulting in specified power to not be obtained was found.There was no patient involvement.
|