| Model Number |
A22041A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: this event caused by a component failure of the ceramic tip (insulation beak).Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the reusable endoscope sheath, to the service center for evaluation related to a separate issue.During testing, the field service engineer identified the device ceramic tip was broken or damaged.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|