| Model Number |
DDMB1D1 |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Tachycardia (2095); Shock from Patient Lead(s) (3162)
|
| Date of Event |
05/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that right ventricular (rv) lead undersensing was observed with no impact to overall device function and far field r-wave (ffrw) oversensing was observed from the right atrial (ra) lead triggering inappropriate mode switch.It was also reported that the implantable cardioverter defibrillator (icd) may have provided inappropriate anti-tachycardia pacing (atp) and high voltage therapy for an supra ventricular tachycardia (svt) episode detected as fast ventricular tachycardia (fvt).Additionally the icd may have delayed therapy for a ventricular tachycardia (vt) episode detected as supra ventricular tachycardia (svt).The ra lead, rv lead and icd remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|