The user reported differences between sg and bg readings.Based on a review of the sensor data available in the data management system (dms), there was some variation in the sensor values, and the user was in the initial post-insertion period ("early wear time"), which is described in the user guide and supported by clinical performance data as an expected phase of sensor stabilization.The user was advised to perform a sensor re-link to allow the system to re-initialize and converge faster.The user was informed that the system is expected to improve following stabilization.The user agreed with this assessment.As per the data management system (dms) review, the system remained in active use with up-to-date information.
|