| Model Number |
CCWET0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient experienced inadequate vision and clinical reason vision was not clear enough for his shooting eye.The iol was exchanged for monofocal non-company replacement lens 07 days following the initial implant procedure.Additional information has been requested but is not available at the time of this report.
|
| |
|
Event or Problem Description
|
|
Additional information received and reported that the iol was exchanged for the same lens model but half a diopter less power indicating the correct lens power was not implanted initially.There is no reported defect or fault with the iol.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information was provided in b.2., b.5., and h.11.Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury/malfunction per regulation.No further reports required.The manufacturer internal reference number is: (b)(4).
|
| |
|