Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
It was reported that the right ventricular (rv) lead exhibited high, out-of-range, undefined pacing impedance and a large number of short v-v intervals.During the replacement of the implantable cardioverter defibrillator (icd) device, the physician observed tissue/debris within the connector port/header area.After connecting the rv lead to a new icd, impedance, sensing, and pacing measurements normalized.The rv lead remains in use.No patient complications have been reported as a result of this event.
|