| Model Number |
DVBB1D4 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
03/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered a lead integrity alert (lia) due to sensing integrity counter (sic) and pacing impedance variation.Oversensing of noise was observed.Tachycardia therapy was programmed off and the rv lead remains in use.It was further reported that the patient was referred to a different facility for rv lead extraction/implant.The physician declined to perform the procedure due to infection on the rv lead.The implantable cardioverter defibrillator (icd) and rv lead remain in use.No further patient complications have been reported as a result of this event.
|
| |
|