| Model Number |
379969 |
| Medical Device Problem Codes |
Ambient Noise Problem (2877); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Double-counted events seen on the rv channel during sensing in a vf episode and high short rv interval count.Impedance is unremarkable and the rv capture control test consistently fails so no data was available regarding pacing threshold.Full lead evaluation with postural testing and isometrics was recommended.Lead remains implanted.
|
| |
|