• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. CDX W/BIBAG BLUE STAR; Dialyzer, High Permeability With Or Without Sealed Dialysate System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. 2008T HD SYS. CDX W/BIBAG BLUE STAR; Dialyzer, High Permeability With Or Without Sealed Dialysate System Back to Search Results
Catalog Number 191126
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction: a1, b7.
 
Event or Problem Description
A biomedical technician (bmt) reported to technical support that a 2008t machine experienced fails multiple self-tests including tmp high soft, act temp high, act temp low, temp high soft, and conductivity high.The bmt stated the temperature was 37.1°c and did not change during the self-test.Customer replaced the actuator-test board.Customer placed the device in self-test twice during the call.The bmt stated the device was now functioning normally and passes the self-test.Upon follow-up, the bmt stated that it was a short electrical issue with the bicarb pump.The bmt was not able to provide more information.No photos were available and no parts were returned for evaluation.The bmt confirmed there was no patient involvement, harm, or adverse events associated with the reported issue.The machine is back in service.
 
Event or Problem Description
A biomedical technician (bmt) reported to technical support that a 2008t machine experienced fails multiple self-tests including tmp high soft, act temp high, act temp low, temp high soft, and conductivity high.The bmt stated the temperature was 37.1°c and did not change during the self-test.Customer replaced the actuator-test board.Customer placed the device in self-test twice during the call.The bmt stated the device was now functioning normally and passes the self-test.Upon follow-up, the bmt stated that it was a short electrical issue with the bicarb pump.The bmt was not able to provide more information.No photos were available and no parts were returned for evaluation.The bmt confirmed there was no patient involvement, harm, or adverse events associated with the reported issue.The machine is back in service.
 
Additional Manufacturer Narrative
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A biomedical technician (bmt) reported to technical support that a 2008t machine experienced fails multiple self-tests including tmp high soft, act temp high, act temp low, temp high soft, and conductivity high.The bmt stated the temperature was 37.1°c and did not change during the self-test.Customer replaced the actuator-test board.Customer placed the device in self-test twice during the call.The bmt stated the device was now functioning normally and passes the self-test.Upon follow-up, the bmt stated that it was a short electrical issue with the bicarb pump.The bmt was not able to provide more information.No photos were available and no parts were returned for evaluation.The bmt confirmed there was no patient involvement, harm, or adverse events associated with the reported issue.The machine is back in service.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).Complaint investigation found objective evidence indicating a product problem, thus the complaint is confirmed.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was able to be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. CDX W/BIBAG BLUE STAR
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key25643990
Report Number0008030665-2026-01483
Device Sequence Number21170460
Product Code KDI
UDI-Device Identifier00840861102099
UDI-Public00840861102099
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K243237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number191126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2026
Supplement Date Received by Manufacturer06/16/2026
07/29/2026
Initial Report FDA Received Date06/26/2026
Supplement Report FDA Received Date06/29/2026
07/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-