Model Number 3401
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 04/19/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The local area field representative was contacted for additional information regarding event resolution.At this time, no further information is available.Should additional information become available this report will be updated.This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Event or Problem Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) and electrode delivered inappropriate shocks due to double- and triple-counting.Surface echocardiogram (s-ecg) review of one treated episode had showed p-wave oversensing, and the morphology did not change post-shock.It was noted that this s-icd was programmed to secondary, 1x gain, and the r-waves were wide and biphasic in the presenting/captured s-ecg.In primary and alternate, the r-waves were narrow.Technical services (ts) recommended considering primary.This electrode remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Common Device Name
Implantable Lead
Manufacturer (Section D)
OSCOR INC
3816 desoto blvd.
palm harbor FL 34683
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key25644747
Report Number2124215-2026-34382
Device Sequence Number18601514
Product Code LWS
UDI-Device Identifier00802526575600
UDI-Public(01)00802526575600(17)170601
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/01/2017
Device Model Number3401
Device Catalogue Number3401
Device Lot NumberA122528
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2026
Supplement Date Received by Manufacturer08/11/2026
Initial Report FDA Received Date06/27/2026
Supplement Report FDA Received Date08/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age57 YR
Patient SexMale
-
-