• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MIDAS REX MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MIDAS REX MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE Back to Search Results
Model Number MR8-AS09-R
Medical Device Problem Codes Vibration (1674); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: evaluation determined that the reported malfunction of loud noise and vibration was confirmed.The likely cause of failure could be damaged bearings.However it was also noted that vague tube and base markings that need etching.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that attachment is not working properly it is having loud noise and vibration.At the time of report, there was no patient involvement.Additional information was received stating that damaged bearings were found during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX MR8
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key25644922
Report Number1625507-2026-03772
Device Sequence Number5902612
Product Code HBE
UDI-Device Identifier00763000063603
UDI-Public(01)00763000063603(11)180814
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AS09-R
Device Catalogue NumberMR8-AS09-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2026
Initial Report FDA Received Date06/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-