| Model Number |
MR8-AS09-R |
| Medical Device Problem Codes |
Vibration (1674); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation determined that the reported malfunction of loud noise and vibration was confirmed.The likely cause of failure could be damaged bearings.However it was also noted that vague tube and base markings that need etching.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that attachment is not working properly it is having loud noise and vibration.At the time of report, there was no patient involvement.Additional information was received stating that damaged bearings were found during evaluation.
|
| |
|