Inappropriate or Unexpected Reset (2959); Compatibility Problem (2960)
Health Effect - Clinical Code
No Clinical Signs, Symptoms or Conditions (4582)
Date of Event
06/19/2026
Type of Reportable Event
Malfunction
Event or Problem Description
Following a magnetic resonance imaging (mri), the device was found to be in backup vvi mode (bvvi) resulting in high-voltage (hv) therapy being disabled.Technical support was contacted, and the device was reprogrammed to resolve the event.The patient was stable with no consequences.