|
ARTHROCARE CORPORATION SUPER MULTIVAC 50 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
ASH4830-01 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
06/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).H11 h3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a synovectomy procedure, a super multivac 50 tip disintegrated and the wand stopped working.The quantum box did not alert, and the settings were set to 9 and 2.A c-arm was taken out to make sure there was not any metal left in the patient.The procedure was resumed, with a 30 minute delay, with a change in surgical technique.The patient experienced post operative severe pain.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|