| Model Number |
DDBC3D4 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Syncope/Fainting (4411); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented due to syncope.A device interrogation indicated the implantable cardioverter defibrillator (icd) treated an episode detected as ventricular fibrillation (vf) that may have been ventricular tachycardia (vt).The right ventricular (rv) lead exhibited undersensing during the episode though there appeared to be no impact to device function.The icd and rv lead remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|