ENCORE MEDICAL L.P. DJO SURGICAL; BASEPLATE, GLENOID HA-COAT, RSP, 6.5MM X 30MM Back to Search Results
Catalog Number 508-32-104
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Laxity (4526)
Date of Event 06/10/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The agent reported: "baseplate central screw and 2/3 peripheral screws broken inside scapular.Baseplate became loose, came off the glenoid.Had to take out original liner to match new 36mm glenosphere." the actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established." item number: 508-32-104, "item description: baseplate, glenoid ha-coat, rsp, 6.5mm x 30mm", lot#: 866c2999, part revision: j, product type: shoulder, manufacture date: 12-may-2022, expiration date: 05-may-2028.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery due to loosening of the baseplate associated with fracture of the central screw and two peripheral screws within the scapula, leading to baseplate dissociation from the glenoid.There are multiple factors that may contribute to an event that are outside of the control of enovis surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.
 
Event or Problem Description
Revision surgery due to baseplate central screw and 2/3 peripheral screws broken inside scapular.Baseplate became loose, came off the glenoid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
BASEPLATE, GLENOID HA-COAT, RSP, 6.5MM X 30MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd.
austin, TX 78758
MDR Report Key25650345
Report Number1644408-2026-01246
Device Sequence Number9750225
Product Code PHX
UDI-Device Identifier00888912024754
UDI-Public00888912024754
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number508-32-104
Device Lot Number866C2999
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2026
Initial Report FDA Received Date06/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
506-03-130 LOT: 835C1836.; 506-03-138 LOT: 837C1174.; 508-00-032 LOT: 855C2363.; 508-32-101 LOT: 862C3633.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
-
-