• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA VAPOR 2000; ANESTHESIA UNIT VAPORIZERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA VAPOR 2000; ANESTHESIA UNIT VAPORIZERS Back to Search Results
Model Number M35054
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation has just started, results will be provided in a follow-up report.
 
Event or Problem Description
It was reported that the anesthetic agent sprayed out of the vaporizer during an operator attempt to fill it up.As per report, the operator was not exposed to the substance.Remark: the particular agent intended for use with this device has a boiling point of 58°c / 136.4°f - it should thus not spray out of the vaporizer at room temperature.At the moment, the manufacturer cannot exclude the presence of a malfunction or use error and thus, this report is filed in abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPOR 2000
Common Device Name
ANESTHESIA UNIT VAPORIZERS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23558
GM   23558
Manufacturer Contact
moislinger allee 53-55
lübeck 23558
4518822868
MDR Report Key25650431
Report Number9611500-2026-00326
Device Sequence Number5877847
Product Code CAD
UDI-Device Identifier04048675228059
UDI-Public(01)04048675228059(11)150205(93)M35054-27
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K971923
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM35054
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/04/2026
Initial Report FDA Received Date06/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/05/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-