Correction: it was determined after receiving further clarification this event for this product was not a reportable event per cfr 21, part 803.This supplemental is being filed to identify an mdr was not required for this event.Additional information: d6a, d6b, h6.The reported infection and device removal were confirmed.The patient developed a draining fistula associated with cutibacterium acnes infection, resulting in removal of the right tmj components and placement of an antibiotic spacer.Medical assessment concluded that the infection was related to bacterial biofilm formation from skin related contamination and not to the tmj device itself.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
|