• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KERI MEDICAL SA TOUCH - CMC1 PROSTHESIS; TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 8MM - M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KERI MEDICAL SA TOUCH - CMC1 PROSTHESIS; TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 8MM - M Back to Search Results
Model Number NTO158
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Metal Related Pathology (4530); Implant Pain (4561); Joint Impingement (4834)
Date of Event 05/11/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information, the incident is not attributed to a manufacturing or design defect of the specific returned device.Mechanical complication and dislocation are known adverse events listed in the device's instruction for use (ifu) and subject to trend reporting.In addition, the explant exhibited signs of initial intra-prosthetic impingement; however, the underlying cause of this impingement could not be determined with certainty.
 
Event or Problem Description
It was reported that a patient underwent a total joint arthroplasty of touch cmc1 prosthesis.Subsequently, the patient presented pain and a dislocated prosthesis, a reduction was performed under fluoroscopy, and malalignment of the neck head was noted therefore, the patient underwent a revision.
 
Additional Manufacturer Narrative
This mdr is being retracted due to misidentification of manufacturing report.A new mdr with the correct manufacturing report has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOUCH - CMC1 PROSTHESIS
Common Device Name
TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 8MM - M
Manufacturer (Section D)
KERI MEDICAL SA
chemin du pre-fleuri 5
plan-les-ouates, 1228
SZ  1228
Manufacturer (Section G)
KERI MEDICAL FRANCE
34 rue antoine redier
archamps, 74160
FR   74160
Manufacturer Contact
camille miletto
chemin du pre-fleuri 5
plan-les-ouates, 1228
SZ   1228
MDR Report Key25651473
Report Number3024706750-2026-00828
Device Sequence Number5902900
Product Code SFA
UDI-Device Identifier07640181160051
UDI-Public07640181160051
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P240020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 06/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberNTO158
Device Catalogue NumberNTO158
Device Lot NumberX5593
Date Returned to Manufacturer06/12/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/29/2026
Supplement Report FDA Received Date06/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-