Event or Problem Description
|
|
Healthcare professional reported "intracapsular rupture".This record is for the left side.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "intracapsular rupture".Later, a healthcare professional reported "rupture," capsular contracture baker grade unknown, "silicone bleed," "leaking silicone," "square appearance," "pain," "hardening," "distortion," and "keyhole sign," and noted that, after a mammogram, the "breast is doing something weird" and the "shape looks off." later, the healthcare professional reported capsular contracture baker grade iii/iv.The patient underwent a capsulectomy.This record is for the right side.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional and/or corrected data: d9, h3, h6.Device evaluation:
per the investigation procedure, the device is analyzed through visual microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.- capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (non-penetrating nick and crease) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|