| Model Number |
VTICMO 13.2 (-15.00/1.00) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault with significant reduction of ica and shallowing of ac.Lens explanted.Problem resolved.Cause of the event was reported as patient-related factor.
|
| |
|
Search Alerts/Recalls
|