| Model Number |
420-PRO-07103000-S |
| Medical Device Problem Code |
Collapse (1099)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The implant remains in situ.Radiographic imaging was provided confirming the event.Review of the device history record (dhr) confirms that implant met all dimensional and inspection requirements at the time of manufacture, with no nonconformances identified.There is no evidence to suggest that manufacturing-related issues contributed to the reported complaint.Based on information provided, a definitive root cause could not be determined.Alphatec spine, inc.Is submitting this report in compliance with fda regulations under 21 cfr 803.All relevant and available information was included to the best of our knowledge at the time of this submission.Any fields left blank reflect information that was not known or not provided.
|
| |
|
Event or Problem Description
|
|
Around 11 months postoperatively, radiographic imagining revealed a collapsed expandable cage.The patient is asymptomatic, and there are no plans for revision surgery.
|
| |
|