• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRAUB MEDICAL AG Rotarex; Thrombectomy & Atherectomy

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRAUB MEDICAL AG Rotarex; Thrombectomy & Atherectomy Back to Search Results
Catalog Number 80236
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Embolism/Embolus (4438)
Date of Event 03/03/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample has not been returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the results of the clinical trial (bdpi-25-001) that during the index procedure, the subject had an adverse event of distal embolization in popliteal bifurcation and distal aspect of the anterior tibial artery.The adverse event was not related to the device but possibly related to the procedure.However, the current status of the patient was unknown.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: no physical sample images videos were received.A physical investigation was not possible.Due to no sample images videos received the reported malfunction cannot be confirmed.Clogging of the catheter appears due to a restricted flow of fluid inside the catheter, that can be lead back to an insufficient dosage of heparin, too fast advance of the catheter or insufficient addition of saline during treatment.It is advised to predilate proximal part of occlusion and to avoid working in the fractured stent area as stent pieces can be aspirated and block the catheter.The insufficient flow can result in the break of the helix and guidewire vessel rapture and other serious damage.A clear root cause could not be identified but a blockage of the catheter represents a known inherent risk.Section a through f : the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the results of the clinical trial (b)(6) that during the index procedure, the subject had an adverse event of distal embolization in popliteal bifurcation and distal aspect of the anterior tibial artery.The adverse event was not related to the device but possibly related to the procedure.However, the current status of the patient was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Rotarex
Common Device Name
Thrombectomy & Atherectomy
Manufacturer (Section D)
STRAUB MEDICAL AG
straubstrasse 12
ch-7323
wangs SG 73 23
SZ  SG 7323
Manufacturer (Section G)
Straub Medical AG
straubstrasse 12
ch-7323
wangs SG 73 23
SZ   SG 7323
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key25662971
Report Number3008439199-2026-00229
Device Sequence Number10447220
Product Code MCW
UDI-Device Identifier07640142811855
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K211738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number80236
Device Lot Number252044
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2026
Supplement Date Received by Manufacturer09/01/2026
Initial Report FDA Received Date06/30/2026
Supplement Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient RaceWhite
-
-