| Catalog Number |
80236 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Embolism/Embolus (4438)
|
| Date of Event |
03/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample has not been returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of the clinical trial (bdpi-25-001) that during the index procedure, the subject had an adverse event of distal embolization in popliteal bifurcation and distal aspect of the anterior tibial artery.The adverse event was not related to the device but possibly related to the procedure.However, the current status of the patient was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: no physical sample images videos were received.A physical investigation was not possible.Due to no sample images videos received the reported malfunction cannot be confirmed.Clogging of the catheter appears due to a restricted flow of fluid inside the catheter, that can be lead back to an insufficient dosage of heparin, too fast advance of the catheter or insufficient addition of saline during treatment.It is advised to predilate proximal part of occlusion and to avoid working in the fractured stent area as stent pieces can be aspirated and block the catheter.The insufficient flow can result in the break of the helix and guidewire vessel rapture and other serious damage.A clear root cause could not be identified but a blockage of the catheter represents a known inherent risk.Section a through f : the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of the clinical trial (b)(6) that during the index procedure, the subject had an adverse event of distal embolization in popliteal bifurcation and distal aspect of the anterior tibial artery.The adverse event was not related to the device but possibly related to the procedure.However, the current status of the patient was unknown.
|
| |
|