• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO, JAPAN; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO, JAPAN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number JP2110X16B
Medical Device Problem Codes Circuit Failure (1089); Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported failure of ac power supply is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review: dhr review was performed for the complaint device serial number, h318013716714 review confirms that the device met the final acceptance criteria and was released for final distribution.
 
Event or Problem Description
The manufacturer received information alleging a trilogy evo was alarming for a service required alarm condition.There was no harm or injury reported.The device was not in patient use.The device was returned to the manufacturer's service center for evaluation and the customer's complaint was confirmed.A service required alarm indicating and ac power failure was observed in the device's downloaded event log.The device's power supply and system board were replaced to address the issue.The device was cleaned and tested and passed all final testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO, JAPAN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25664129
Report Number2518422-2026-027880
Device Sequence Number13356929
Product Code CBK
UDI-Device Identifier00606959051973
UDI-Public(01)00606959051973
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberJP2110X16B
Device Catalogue NumberJP2110X16B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2026
Initial Report FDA Received Date06/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/02/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-