ICU MEDICAL, INC. CADD-SOLIS VIP AMBULATORY INFUSION PUMP; PUMP, INFUSION Back to Search Results
Model Number 2120
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event or Problem Description
It was reported that the cadd-solis vip ambulatory infusion pump had a ripped and bubbled downstream occlusion (dso) seal, and the air sensor was found to be detached and falling off.The issue was identified during testing.There were no patient involvement and no patient harm.This may lead to fluid ingress into the pump which will interrupt therapy during infusion if it recurs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS VIP AMBULATORY INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ICU MEDICAL, INC.
3350 granada avenue north
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
+1(224)706-2300
MDR Report Key25664788
Report Number3012307300-2026-06664
Device Sequence Number5596711
Product Code FRN
UDI-Device Identifier15019517096378
UDI-Public(01)15019517096378(11)171117
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K111275
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2120
Device Catalogue Number21-2120-0103-51
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/25/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2026
Supplement Date Received by Manufacturer08/19/2026
Initial Report FDA Received Date06/30/2026
Supplement Report FDA Received Date09/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2017
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2419-2024, Z-2421-2024
Patient Sequence Number1
Patient SexUnknown
-
-