Catalog Number G122H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used.It was reported that the tip of the suture breaks off when suturing.Recurring; happens every single use (clinic uses about 30 per week).Product is not modified and is used correctly according to vendor instructions.The clinic has never had this problem before with this product ("something has changed").Clinic is accounting for the broken tips, using additional sutures to complete procedures, and is actively sourcing new sutures.Doctor would also like to add he has been practicing surgery for more than 25 years, both in his private sedation clinic and at the northern regional hospital, and has never encountered this type of problem with any sutures during his career.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 6/30/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The following information was received: a request for a return kit was requested just a couple days ago, as we were not told of the specifics in regards to sending them back.The product is available and on its way back from the customer, but then we will require a prepaid shipping label be issued to send the boxes back.Additional information was requested, the following was obtained: how many procedures/patients were affected by this issue? we do 28 surgeries a week i would say at least 1/3 of those.How many devices demonstrated the alleged deficiency in each patient event? if possible, please specify how many devices were associated with lot 10be7q, how many with lot 10aeqd, and how many with lot 10bh9k.I don¿t have an answer for this the best i can say is each surgery we use 1 suture and don¿t track the lot numbers of what we use on each pt.Lot 109386 this should be 1093b6 was also provided; however, this lot number is valid for a surgicel/mesh product.Please confirm whether this is the correct lot number.The event description states that the tip of the needle breaks off during suturing; however, the photos provided show bent needles.Could you please clarify whether the needle tip broke off as a separate piece, or whether the needle bent without breaking? we took the photos before they broke off.We have had the tip break off and be in the pt¿s mouth i will take more photos.Were there any patient consequences associated with any of these events? non that we are aware of at this time please provide the title of external person providing answers to follow-up (e.G.Nurse, surgeon, risk manager (b)(6) licensed surgical dental assistant on behalf of dr.(b)(6) oral and maxillofacial surgeon.Please provide the status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details i have them all packaged and ready to go if there are further details requested, please let us know.As well, kindly confirm whether a shipping label will be provided for the return of the items for investigation.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHICON
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key25665103
Report Number2210968-2026-06884
Device Sequence Number5765053
Product Code GAL
UDI-Device Identifier10705031037441
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Type of Report Initial,Followup
Report Date (Section B) 06/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberG122H
Device Lot Number10BH9K
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2026
Supplement Date Received by Manufacturer07/07/2026
Initial Report FDA Received Date06/30/2026
Supplement Report FDA Received Date07/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/13/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-