• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number CAX500T12C
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected products with field complaints alleging pe-pur foam degradation have undergone the establishment of complaint codes specifically identifying foam degradation, for dream station cpap/bipap series.A field correction action(2021-06-a) was initiated, and medical device recall letters were issued.A field action was initiated to replace pe-pur foam in affected products; degradation was added to the respective dfmeas and the risk management file that was current at the time ¿ er 2219697 v15 (nirvana risk management matrix) was updated to include hazards associated with foam degradation based on complaints analyzed through various health hazard evaluations (hhes).Risk tags ¿ degrad04, irrit05 reflected the safety risk assessment of design containing the affected pe-pur foam.These complaints were monitored and trended in accordance with the complaint trending procedures.If the complaints were determined to meet the criteria for escalation, they were escalated for risk assessment through the post market clinical escalation process.As of the date of submission for this report, there is no indication that complaints for the failure in question have led to unacceptable levels of severity or probabilities of harm, and therefore no further action will be pursued.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
 
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.A dreamstation auto cpap device was returned to a third-party service center with no initial alleged complaint.During the evaluation of the device at the third-party service center, the device was visually inspected, and evidence of visible foam particles/ degradation was found inside the blower kit.Additionally, the service technician also noted dust contamination.The device was scrapped by the service center after the evaluation was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25668830
Report Number2518422-2026-028092
Device Sequence Number10447365
Product Code BZD
UDI-Device Identifier00606959034549
UDI-Public(01)00606959034549
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 06/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCAX500T12C
Device Catalogue NumberCAX500T12C
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2026
Initial Report FDA Received Date06/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/24/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-