Model Number DSX500T11
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2026
Type of Reportable Event Malfunction
Event or Problem Description
A dreamstation auto cpap device was returned to a third-party service center with no initial alleged complaint.There was no report of serious or permanent harm or injury, and no medical intervention was reported.During evaluation at the third-party service center, the device was visually inspected and evidence of foam degradation/particles was observed inside the blower kit.Blower foam degradation (blfm) was also identified.Additionally, the service technician noted dust contamination.Following completion of the evaluation, the device was scrapped by the service center.There was no reported patient involvement, injury, or adverse clinical outcome associated with this event.Based on the identification of foam degradation/particles and blower foam degradation during device evaluation, this event was determined to be reportable under fda 21 cfr part 803 as a product malfunction and/or potential serious injury event.No additional information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25668862
Report Number2518422-2026-028156
Device Sequence Number16053909
Product Code BZD
UDI-Device Identifier00606959424197
UDI-Public00606959424197
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 06/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDSX500T11
Device Catalogue NumberDSX500T11
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2026
Initial Report FDA Received Date06/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/02/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1972-2021
Patient Sequence Number1
Patient SexUnknown
-
-