|
A dreamstation auto cpap device was returned to a third-party service center for evaluation.During evaluation at the third-party service center, the device was visually inspected and evidence of foam particles was observed inside the blower kit.Additionally, the service technician noted dust contamination.Following completion of the evaluation, the device was scrapped by the service center.There was no report of death, serious or permanent harm, injury, or medical intervention associated with this event.Based on the identification of foam particles inside the blower kit during device evaluation, this event was determined to be reportable under fda 21 cfr part 803 as a product malfunction and/or potential serious injury event.No additional information is available at this time.
|