SYNTHES GMBH Q-COUPLING FOR KWIRES AND PINSDIAMETER1.5-4.0MM; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.791
Medical Device Problem Codes Mechanical Jam (2983); Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during service and evaluation, it was determined that the q-coupling for kwires and pinsdiameter1.5-4.0mm device was frozen/would not move.It was noted in the service order that the device could be grasped but would not turn.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: check cannulation of attachment.Check function in running mode.During the service evaluation the following failures were identified: functional : frozen/will not move.Visual : cosmetic damage.Conclusion: failure reported is confirmed.Root cause: faulty parts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Q-COUPLING FOR KWIRES AND PINSDIAMETER1.5-4.0MM
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
elba bello
eimattstrasse 3
oberdorf 04436
SZ   04436
9083863534
MDR Report Key25669771
Report Number8030965-2026-06751
Device Sequence Number10407041
Product Code GEY
UDI-Device Identifier07611819491861
UDI-Public(01)07611819491861
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number530.791
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2026
Initial Report FDA Received Date06/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/15/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-