The agent reported "the adverse event led to the empowr 3d liner revision due to periprosthetic joint infection" the previous surgery and the surgery detailed in this event occurred 46 days apart.Item number: 341-10-709, item description: empowr 3d kneetm ins, 9l 10mm, ve, lot number: 069t1251, part revision: f, product type: knee, manufacture date: 1/14/2026, expiration date: 1/9/2031, this evaluation is limited in scope as limited information was provided to djo surgical - austin for examination.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was the source or had a direct connection with the patient's infection.Root cause: the root cause of this complaint was reported as an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post surgical instructions.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history record(s) show that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or on-going issues that are needing a review.This is the first complaint against this item/lot number combination across all failure code categories.There are no indications that this item has a design or material deficiency.Sales for past two years: 27.No similar incidents in germany in part two years.
|