BOSTON SCIENTIFIC CORPORATION AMS 700 LGX; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72404292
Medical Device Problem Codes Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with an inflatable penile prosthesis (ipp) presented with a nonfunctioning device.A revision surgery was performed, during which the physician found that the cylinder had a tear.The device was explanted and replaced.There were no patient complications.
 
Additional Manufacturer Narrative
Model number/catalog number: 72404310 serial number: n/a batch/lot number: 0144803001 model/catalog description: pump ms unique identifier (udi) #: (b)(4).Model number/catalog number: 72404161 serial number: n/a batch/lot number: 0144729017 model/catalog description: reservoir 65ml pc unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Mechanical issue and hole/perforated are acknowledged within the instructions for use (ifu) and are not related to off-label use or failure to follow instructions.Based on the information available the cause that contributed to the reported event cannot be established; therefore, a conclusion code of cause not established was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 LGX
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key25673418
Report Number2124215-2026-34853
Device Sequence Number21445712
Product Code JCW
UDI-Device Identifier00878953003948
UDI-Public(01)00878953003948(17)210520(10)135687004
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/20/2021
Device Model Number72404292
Device Catalogue Number72404292
Device Lot Number0135687004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2026
Initial Report FDA Received Date06/30/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age76 YR
Patient SexMale
-
-