BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER Back to Search Results
Model Number 3300
Medical Device Problem Code Key or Button Unresponsive/not Working (4063)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the integrated programmer screen froze while saving to universal serial bus (usb).Boston scientific technical services (ts) suggested to power down and re-interrogate.Subsequently, the data was successfully saved to a drive, not the usb.As of this time, this programmer remains in service.No patient involvement.
 
Additional Manufacturer Narrative
The device was not returned for product analysis.Boston scientific performed an investigation with the available information.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.Investigation determined that the most common cause of the touchscreen becoming unresponsive during use was the water detection feature within the programmer liquid crystal display (lcd) touchscreen assembly.The purpose of this water detection feature is to detect the presence of water on/near the programmer screen.If water is detected, this feature is designed to disable the touchscreen to prevent water droplets from causing false positive touch inputs during use.Specifically, the investigation determined that various inputs could induce false positive detections of this water detection feature, which then results in an unresponsive touchscreen.Our investigation also identified additional potential causes independent of the water detection feature.These include a variety of unrelated root causes such as hardware, environmental, software, and user-related causes that can be intermittent and not reproducible during laboratory analysis.A risk review was completed and confirmed that the event of unresponsive was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause traced to device design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE PROGRAMMING SYSTEM
Common Device Name
EXTERNAL PROGRAMMER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
PLEXUS CORPORATION
2400 millbrook dr
buffalo grove IL 60089
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key25674408
Report Number2124215-2026-35006
Device Sequence Number15867217
Product Code OSR
UDI-Device Identifier00802526608513
UDI-Public(01)00802526608513
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910077/S159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number3300
Device Catalogue Number3300
Device Lot Number008427
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2026
Initial Report FDA Received Date07/01/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-