• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1054097
Medical Device Problem Codes Failure to Charge (1085); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to a third-party service center for service.There was no harm or injury reported.During the evaluation of the device at third-party service center, a "service required" code was found in the ventilator's downloaded error log.The device's internal battery was replaced, and the software was upgraded to address the issue.Additionally, the patient¿s complaint was not confirmed along with contamination, and the device was corrected.Ac power cord retainer is missing.Rubber feet are missing.Handle is cracked.Front logo is deteriorated.
 
Additional Manufacturer Narrative
The reported failure of a "service required" code is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key25675651
Report Number2518422-2026-114307
Device Sequence Number18574725
Product Code CBK
UDI-Device Identifier00606959022751
UDI-Public(01)00606959022751
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1054097
Device Catalogue Number1054097
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2024
Initial Report FDA Received Date07/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/18/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-