| Model Number |
Mi1250 SYNCHRONY 2 PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
Active electrodes of the device have been affected.
|
| |
|
Event or Problem Description
|
|
In-situ measurements show an increase in channels affected by high impedances since may 2023.Due to complex medical history, re-implantation is not recommended; the audiologist is re-programming the right processor for sound awareness, disabling the hi electrodes and electrodes showing significant fluctuation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.Mechanical influences, like the reported head hitting may have led to a weakening of the fixation and therefore may have led to device mobility.However, to determine an exact root cause a device investigation of the explanted device is necessary.Due to the complex medical history the recipient will not undergo revision surgery.Affected channels have been disabled and recipient will be further monitored.
|
| |
|
Search Alerts/Recalls
|