H3: product analysis :evaluation determined that device was seized up , and can't move.The likely cause was identified as due to input drive shaft which is stuck and cannot rotate.It was also noted that distal bearing is detached , input and output drive shaft and tactile o-ring sleeve are worn , and also laser markings and color ring are faded, bearings were also found to be corroded.B.3.Date of incident or estimated date of incident was not provided to the manufacturer.Event notified date used as date of incident until further information is obtained.G.3.Aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|