| Catalog Number |
863017D |
| Medical Device Problem Codes |
Premature Activation (1484); Self-Activation or Keying (1557)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a breast tissue marker placement procedure using the ultra clip dual trigger.It was reported that prior to the procedure, just after opening the package device was excited and unusable.The procedure was completed using another device.There was no patient contact.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a surgeon was preparing to use breast tissue marking cannula on patient using ultra clip dual trigger.Upon opening the package, the surgeon discovered that the cannula had been fired and was unusable.A new breast tissue marking cannula was immediately substituted, and the incident was reported as an adverse event.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr and supplemental report 1 listed an incorrect manufacturing location (g1).The correct manufacturing location for this device is bard shannon limited.No new event information is being reported in this supplement.Based on the sme assessment and review of prior similar events, the reported premature activation issue was determined to be not reportable.As the premature activation code is not specifically addressed in the reportability guidelines and was assessed by the sme as non-reportable, the event no longer met the criteria for mdr reportability.Therefore, the complaint was downgraded to not reportable.However, because an initial mdr had already been submitted, the complaint remains classified as a malfunction, and a supplemental mdr will be submitted to document the updated reportability assessment and the rationale for the downgrade.Manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported premature activation could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G1.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|