| Model Number |
VT2110X24B |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of the trilogy evo machine flow sensor is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Review of the dhr for this device, serial number (b)(6), did not find any non-conformances or abnormalities related to the reportable malfunction/ allegation identified in this complaint record during testing of the device prior to distribution.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information that a trilogy evo ventilator was returned to a third-party service center for evaluation.There was no patient involvement, and no harm or injury reported.On evaluation, ventilator service required alarms e-384 and e-386 indicating the device software has detected a large pressure drop between the monitor and outlet pressure sensors (i.E.Across the machine flow sensor).The machine flow sensor required replacement to address the issue.The device then passed functional testing.
|
| |
|
Search Alerts/Recalls
|