Model Number 407652
Medical Device Problem Codes Pacing Problem (1439); Under-Sensing (1661); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient experienced syncope.The right ventricular (rv) lead exhibited high and rising thresholds and the right atrial (ra) lead exhibited out of range low impedance and undersensing during atrial tachycardia/atrial fibrillation (at/af) st ored episodes resulting in termination of at/af detected events in progress, lack of detection of at/af, and inappropriate pacing.It was also noted that there were low amplitude p waves.Both leads remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 407458 lead implanted (b)(6) 2016.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS MRI SURESCAN
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key25678359
Report Number2649622-2026-19510
Device Sequence Number13175877
Product Code DTB
UDI-Device Identifier00643169550810
UDI-Public(01)00643169550810(17)171029
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/29/2017
Device Model Number407652
Device Catalogue Number407652
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2026
Initial Report FDA Received Date07/01/2026
Date Device Manufactured11/11/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
W1DR01 IPG
Patient SexMale
Patient Weight75 KG
-
-