| Model Number |
479878 |
| Medical Device Problem Code |
High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: dtpa2qq crtd - implanted:(b)(6) 2023; 6935m55 lead - implanted: (b)(6) 2023; 479878 lead - implanted: (b)(6) 2023.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported by the patient that they believe the battery "should last longer" on their cardiac resynchronization therapy defibrillator (crt-d).It was further reported that the left ventricular (lv) lead exhibited high threshold contributing to the battery depletion and reprogramming was performed and the battery depletion currently seems to be consistent with current lv output threshold.The lv and the crt-d remains in use.No patient complications have been reported as a result of this event.
|
| |
|