|
RESPIRONICS, INC. BIPAP A40 PRO; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
| Model Number |
DEX3100S13 |
| Medical Device Problem Codes |
Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of ventilator inoperative is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.A field action was initiated with record id 2023-cc-src-039 and consisted of a field safety notice.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review was performed for the complaint device serial number (b)(6) review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information that a bipap a40 pro device was returned to a third-party service center for service.During evaluation of the device, the event log was reviewed and there was a ventilator inoperative error code, e50, in the device's history.This issue has been identified under the fsn 2023-cc-src-039 due to interruptions and/or loss of therapy due to a ventilation inoperative alarm.There was no patient harm or serious injury reported with this notification.The device was not in patient use.During the evaluation of the bipap a40 pro device at a third-party service center, the event log was reviewed and there was a ventilator inoperative error code, e50, indicating that the difference between the blower pressure sensor reading and the outlet pressure sensor reading is 10 cmh2o or greater for 5 seconds.The technician recommended replacing the device's printed circuit assembly (pca) to address the issue.Additionally, the evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.No repair was performed.The device was scrapped by the service center after the evaluation was completed.
|
| |
|
Search Alerts/Recalls
|
|
|