| Catalog Number |
530.705 |
| Medical Device Problem Codes |
Physical Resistance/Sticking (4012); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the battery handpiece device had a sticky trigger.It was noted in the service order that the device did not work.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A visual and functional assessment was performed on the device.The following steps failed: section 70: check for sticky trigger.Section 80: check function of device.Section 120: check power with power test bench.During the service evaluation the following failures were identified: functional: will not run - electrical control unit damaged.Functional: sticky trigger.Conclusion: failure reported is confirmed.Root cause: component failure from wear.
|
| |
|
Search Alerts/Recalls
|