Model Number NVRO-35
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2026, a 35mm navitor vision valve was selected for implant using a large flexnav delivery system (ds).During the procedure, on 80 percent of the deployment, an incomplete stent opening (iso) was observed and a recapture was attempted to redeploy.However, the valve was unable to be recaptured completely; micro-adjustment wheel was used when the ds was pulled at the descending aorta, but the gap remained unchanged.The ds with the half-captured valve was removed from the patient's anatomy through a 20fr unspecified sheath.A new 35mm, navitor vision valve was selected for implant using a new large flexnav ds/ during the procedure, the patient became hypotensive.The user believes the valve expanded non-uniformly due to the excess calcium.The patient was discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVITOR VISION
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key25680876
Report Number2135147-2026-04563
Device Sequence Number4707190
Product Code NPT
UDI-Device Identifier05415067045768
UDI-Public(01)05415067045768(17)280616
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
P190023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNVRO-35
Device Catalogue NumberNVRO-35
Device Lot Number10940542
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2026
Initial Report FDA Received Date07/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-