| Model Number |
104523 |
| Medical Device Problem Codes |
High impedance (1291); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Lead information: product description: evoke cap12 percutaneous lead kit - 90cm model # : 102879 catalog#: 3009 serial/lot #: (b)(6).
|
| |
|
Event or Problem Description
|
|
A patient implanted with an evoke spinal cord stimulation (scs) system underwent an impedance check during the initial programming visit, and disconnected electrode and high impedances were identified.An x-ray revealed a lead migration.The patient reported difficulty adhering to post-procedural restrictions.A revision procedure was performed, during which the lead and anchor were replaced.
|
| |
|