• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 620-2211511-01
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.During investigation of the device to determine root cause, the technician observed physical damages to the lcd consistent with mis-handling.The lcd was replaced.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during functional testing, the device's screen turned completely white and was illegible.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key25681803
Report Number1220908-2026-02271
Device Sequence Number4656882
Product Code MKJ
UDI-Device Identifier00847946019075
UDI-Public(01)00847946019075
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number620-2211511-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2026
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2026
Initial Report FDA Received Date07/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-