U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA; Cochlear Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA; Cochlear Implant Back to Search Results
Model Number Mi1260 SONATA 2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 06/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The implant had to be removed because the of a positive culture at the implant site.
 
Additional Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The implant had to be removed because the of a positive culture at the implant site.
 
Additional Manufacturer Narrative
Additional information: according to information received from the field, the recipient had surgical wound dehiscence, with pseudomonas being isolated from the fat tissue inside the cavity.The user reportedly has diabetes and poor glucose control.It was also reported that the device was working properly and that the patient was satisfied.A review of the device sterilization records shows that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the reported issue.The device has been explanted but has not been received yet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONATA
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key25689004
Report Number9710014-2026-00616
Device Sequence Number9731579
Product Code MCM
UDI-Device Identifier09008737385385
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMi1260 SONATA 2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2026
Supplement Date Received by Manufacturer06/29/2026
Initial Report FDA Received Date07/02/2026
Supplement Report FDA Received Date09/22/2026
Date Device Manufactured07/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexMale
-
-