|
Block h6: imrdf code a15 captures the reportable event of clip difficult to release from catheter.Block b3: approximated based on the date the manufacturer became aware of the event.Block h11: investigation results - the mantis was implanted and not returned for analysis; therefore, a technical analysis could not be performed.Product analysis could not be carried out, for this reason and due to the lack of evidence is unable to confirm the reported event.A review of the device history record (dhr) was performed.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on the event description and information provided by the customer there is not enough evidence to determine whether the issue was induced due to the physician's technique during the procedure or was related to a device malfunction.Additionally, based on the most relevant information for the complaint including product record review results, the device met all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.The definitive cause of the reported issue cannot be established due to lack of evidence.Without proper evaluation of the device, it remains unknown the causes that contributed to the event.Based on the thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|