Catalog Number 0500325E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 05/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A dialysis manager reported to fresenius that a dialyzer blood leak occurred during a patient¿s hemodialysis (hd) treatment.Additional information was obtained through follow-up.The blood leak occurred a little over one hour into the treatment.The blood leak was reported to be internal, and it was confirmed to be visually observed.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.A blood leak test strip was not used.There was no damage or defect noted on the dialyzer.Fresenius bloodlines were also being used.After the leak occurred, the lines were clamped off and the treatment was stopped.The patient¿s blood was not returned; estimated blood loss (ebl) was 200 to 250 ml.It was confirmed there was no patient serious injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed their treatment after being re-setup with new supplies on a different machine.The sample was not available to be returned for manufacturer evaluation as it was reportedly discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX F250NRe HF 12/CS 2.5SA EBEAM
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key25691375
Report Number0001713747-2026-00261
Device Sequence Number18504917
Product Code KDI
UDI-Device Identifier00840861100170
UDI-Public00840861100170
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K162488
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0500325E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2026
Supplement Date Received by Manufacturer07/20/2026
Initial Report FDA Received Date07/02/2026
Supplement Report FDA Received Date07/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 2008T MACHINE; FRESENIUS BLOODLINES
Patient SexUnknown
-
-