• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK RESILIENCE 55; LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK RESILIENCE 55; LEAD Back to Search Results
Model Number 457852
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 06/23/2026
Type of Reportable Event Serious Injury
Event or Problem Description
System was explanted due to inadequate pain relief.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESILIENCE 55
Common Device Name
LEAD
Manufacturer (Section D)
BIOTRONIK
6024 jean road
lake oswego OR 97035
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key25691430
Report Number3027063190-2026-00255
Device Sequence Number13122516
Product Code LGW
UDI-Device Identifier04035479170332
UDI-Public(01)04035479170332(17)270630
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P210037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number457852
Device Catalogue Number457852
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2026
Supplement Date Received by Manufacturer07/06/2026
Initial Report FDA Received Date07/02/2026
Supplement Report FDA Received Date07/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
Patient SexUnknown
-
-