| Model Number |
GIF-1TH190 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.In addition, device evaluation found foreign material inside the instrument channel based on the results of the investigation, the most likely cause of the foreign material in the scope was unable to be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported that the gastrointestinal videoscope had black fluid coming out of the distal tip.The issue occurred after reprocessing while the scope was hung up to dry.There was no patient involvement.
|
| |
|